If the Always perform maximum determinations radio button is selected,
the last flow times will be used for the viscosity determination.
When you have completed the configuration and saved the current
settings, you may wish to permanently store the settings using the Save
Instrument and Tray Settings.Tray Settings: Wash
Use the Tray Settings: Wash window to determine wash settings. For
multi-tube units, different wash settings may be selected for each tube
individually.
The Tray Settings: Wash window (Dual Solvent options shown)
Solvent rinse settings
The solvent rinse settings for Solvent A and (if applicable) Solvent B are
determined using the Tray Settings: Wash window. Select a value from
0 to 18 cycles for this option (the “0” setting turns the solvent option off).
These settings determine the number of rinse cycles performed by the
miniAV-LT instrument after each test. Higher viscosity materials may
require more rinses. Lower viscosity materials may require fewer rinses,
depending on the effectiveness of the solvent.